LINEARITY of a DENSITOMETER

Purpose of the test

This test is performed to determine the linearityof a (spectro)densitometer by optical density, using a (spectro)densitometer, of a printed bar on paper.

Procedure

Operate the instrument in accordance with ISO 5-4 and the manufacturer’s instructions.

Leave the samples exposed for 24 h to 23°C and 50% R.H., according to ISO 187.

Adjust the (spectro)densitometer to the following settings: A-illuminant, Status E, polarizing.

Place the test piece in the position indicated on the cover sheet on a black backing.

Without touching the test area use the procedure appropriate to the instrument to obtain the optical densities.

Report 2 measures of optical density per level, their mean and standard deviation calculated to 3 significant figures.

ISO 5-4
Nr. of samples

A pack made of a coloured cover and 2 printed papers

Each printed paper contains 2 strips with colour bars

No special size

Levels

Level 1: White

Level 2: Black 40%

Level 3: Black 80%

Level 4: Black 100%

Density:

Level 1: -0.01÷0.03

Level 2: 0.3÷0.4

Level 3: 0.5÷1.5

Level 4: 1.0÷2.5

Reporting

Report 2 measures of Optical Density per field, their mean and their standard deviation. All to three significant digits.

CEPI test 9.7